Device Identification:
device_id
-
External link: Device identification from CXR algorithm (Cardia X)
-
External link: Device identification with neural networks
-
External link: Normal and abnormal radiographic appearances of implantable cardiac electronic devices
PBLSTOP method for device interrogation
- Presenting rhythm: Paced/sensed/dependent
- Battery status: Voltage, check ERI voltage limit => charge time and date
- Lead status (Atrial, ventricular and voltage impedances)
- Sensing: _ mV
- Threshold: a. Atrial b. Ventricular
- Observations/Events/Diagnostics
- Programming: Make programming changes as necessary and Print
Cardiac Implantable Device Interrogation - Note Template
Device : ***
Battery status: Satisfactory Voltage: *** V
Underlying rhythm: ***
Leads:
RA
RV
CS
RV Coil
SVC Coil
Sensing
mV
mV
mV
Impedance
ohms
ohms
ohms
ohms
ohms
Threshold
V @ 0.4 ms
V @ 0.4 ms
V @ 0.4 ms
Mode: ***
Lower rate/UTR: 60/130 ***
Pulse amplitude A/V: 2.5/2.5 ***
Pulse width A/V: 0.4/0.4 ***
VT1/VT2/VF (bpm): 171/200/240 ***
Last charge time: 8.3 s ***
AS-VS:, AS-VP:, AP-VS:, AP-VP:
Episodes: ***
Changes made: None
Conclusion: Normal device function
Perioperative ICD/Pacemaker Management – Note Template
• ICD detection may be suspended by magnet as needed
• External defibrillation equipment should be immediately available in the procedure area
• Patient should be on a cardiac monitor through procedure and recovery
• Recommend continuous blood pressure monitoring
• Because of interactions with monitoring, ventilation, and other impedance monitoring operative devices, consider inactivating minute ventilation sensors
External link: Device Interrogation and Programming Workshop (HRS 2013)
- Presenting rhythm: Paced/sensed/dependent
- Battery status: Voltage, check ERI voltage limit => charge time and date
- Lead status (Atrial, ventricular and voltage impedances)
- Sensing: _ mV
- Threshold: a. Atrial b. Ventricular
- Observations/Events/Diagnostics
- Programming: Make programming changes as necessary and Print
Cardiac Implantable Device Interrogation - Note Template
Device : ***
Battery status: Satisfactory Voltage: *** V
Underlying rhythm: ***
Leads:
RA |
RV |
CS |
RV Coil |
SVC Coil |
|
Sensing |
mV |
mV |
mV |
||
Impedance |
ohms |
ohms |
ohms |
ohms |
ohms |
Threshold |
V @ 0.4 ms |
V @ 0.4 ms |
V @ 0.4 ms |
Mode: ***
Lower rate/UTR: 60/130 ***
Pulse amplitude A/V: 2.5/2.5 ***
Pulse width A/V: 0.4/0.4 ***
VT1/VT2/VF (bpm): 171/200/240 ***
Last charge time: 8.3 s ***
AS-VS:, AS-VP:, AP-VS:, AP-VP:
Episodes: ***
Changes made: None
Conclusion: Normal device function
Perioperative ICD/Pacemaker Management – Note Template
• ICD detection may be suspended by magnet as needed
• External defibrillation equipment should be immediately available in the procedure area
• Patient should be on a cardiac monitor through procedure and recovery
• Recommend continuous blood pressure monitoring
• Because of interactions with monitoring, ventilation, and other impedance monitoring operative devices, consider inactivating minute ventilation sensors