Device : ***
Battery status: Satisfactory Voltage: *** V
Underlying rhythm: ***
Leads:
|
RA
|
RV
|
CS
|
RV Coil
|
SVC Coil
|
Sensing
|
mV
|
mV
|
mV
|
|
|
Impedance
|
ohms
|
ohms
|
ohms
|
ohms
|
ohms
|
Threshold
|
V @ 0.4 ms
|
V @ 0.4 ms
|
V @ 0.4 ms
|
|
|
Mode: ***
Lower rate/UTR: 60/130 ***
Pulse amplitude A/V: 2.5/2.5 ***
Pulse width A/V: 0.4/0.4 ***
VT1/VT2/VF (bpm): 171/200/240 ***
Last charge time: 8.3 s ***
AS-VS:, AS-VP:, AP-VS:, AP-VP:
Episodes: ***
Changes made: None
Conclusion: Normal device function
• ICD detection may be suspended by magnet as needed
• External defibrillation equipment should be immediately available in the procedure area
• Patient should be on a cardiac monitor through procedure and recovery
• Recommend continuous blood pressure monitoring
• Because of interactions with monitoring, ventilation, and other impedance monitoring operative devices, consider inactivating minute ventilation sensors